Replacing The Pilot Valve Manifold; Reinstalling The Electronic Module - Ametek 5800 User Manual

Moisture
Table of Contents

Advertisement

collar and pulling the tubing free. Make sure that you identify the tub-
ing for proper position when replacing.

Replacing the pilot valve manifold

1.
Position the pilot valve manifold on the electronic module mounting
plate. Attach with the two screws you removed earlier.
2.
Attach the tubing you identified and removed earlier. Make sure that
you are attaching the correct tubing to each valve.
3.
Reattach the pilot valve manifold assembly to the sample manifold
plate using the hex-head captured screws.

Reinstalling the Electronic Module

Once you have serviced the various components in the electronics mod-
ule, reassemble the analyzer.
1.
Carefully reposition the electronic module above and to the front of
the analyzer. Lower the module into the analyzer so the screws at the
four corners can be attached.
2.
Attach the mounting plate of the electronic module to the analyzer as-
sembly using the four screws you removed earlier.
|