Installing The New Solenoid - Ametek 5800 User Manual

Moisture
Table of Contents

Advertisement

Figure 7-7b.
Model 5830 solenoid
replacement.
6.
Using a flat-head screwdriver, remove the screws that hold the valve
body to the manifold; set the screws aside. Lift-off the solenoid valve
and reserve the O-rings for the replacement solenoid.

Installing the new solenoid

1.
Carefully, mount the new solenoid valve on the valve manifold using
the two flat-head screws. Ensure that both O-ring seals are properly
seated.
2.
If you are replacing one of the three-way valves, reconnect the 1/16
inch fitting to the appropriate port on the exhaust manifold.
3.
Place the sample manifold plate back in the oven and secure with the
four mounting screws.
4.
Reconnect all four 1/16 fittings on the bracket in the center of the oven
(Figure 7-2).
5.
Replace the oven lid and secure it with the two nuts (Figure 7-1).
6.
Replace the analyzer cover and secure it with the mounting screws.
Proportional
Solenoid Valve
Electrical
Connection
6/32 Flat Head
Screw (2)
3-Way
Solenoid Valve
2-Way
Solenoid Valve
O-Ring Seal (2)
3-Way
Solenoid Valve
Flat Washer (2)
Lock Washer (2)
3-Way
Solenoid Valve
|
Service & Parts
7-9

Advertisement

Table of Contents
loading