Verify; Abort; Alarm Key - Ametek 5800 User Manual

Moisture
Table of Contents

Advertisement

Alarm Key

Figure 4-4.
Main menu for Alarm
key.
4-4 | 5830 Moisture Analyzer

Verify

This function initiates a verification cycle. Note that a zero cycle is auto-
matically performed, prior to running the internal moisture generator.

Abort

This function aborts a zero/verify cycle and returns the analyzer to normal
operation.
Use the Alarm Key to define alarm output settings.
ALARM
High Limit
Enter the high limit value in moisture units selected.
Low Limit
Enter the low limit value in moisture units selected.
Enable Alarm
Select Enable or Disable to activate the moisture concentration alarm
relay.
High Limit
Low Limit
Enable Alarm
Enter Value
Enter Value
Enable
Disable