Electrical Connections (Figure 3-6)

WARNING

Voltage selection

The 5830 analyzer comes from the factory set to operate on 120 VAC. If
you must switch the setting to 240 VAC (or back to the 120 VAC setting),
follow the procedure below.
CAUTION
1.
2.
3.
4.
5.
Figure 3-6.
115 / 230 voltage
selection switch.
This analyzer has an internal AC voltage selector that must be set
!
correctly to prevent damage. Only qualified service personnel may ac-
cess the interior of the analyzer. If qualified service personnel are not
available, contact AMETEK service.
You must follow this procedure to prevent damage to/malfunction of
!
the analyzer.
Ensure that the area is non-hazardous and the analyzer has been de-
energized and power disconnected
Remove the terminal cover on back panel.
Select 115 or 230 VAC on the voltage selector (10) located on the top
left-hand side of the user interface board.
Replace terminal cover.
Clearly mark the operating voltage on the rear panel analyzer serial
number label in the lower right-hand corner.
LINE VOLTAGE
10
Voltage
Selection
FUSE
Switch
L
N
1
2
TB1
TB2
DATA
CONC
VALID
ALARM
G
3
1
2
3
4
Installation & Start-Up | 3-11
RS-485
TERMINATOR
ON
TB3
ANALOG
OUTPUT
SYSTEM
RD
TD
ALARM
-
+
+
-
+
-
OUT
IN
5
6
7
8
9
10
11
12
13 14
ON