5-4 | 5830 Moisture Analyzer

Echo (A)

Echo implements a loop back function to test the integrity of the serial
communication. Any text transmitted by the master is returned.
:
Example
Master:
>99ABCDEFG4E[CR]
Slave:
AABCDEFG1D[CR]

Bad Command (B)

This command is not recognized by the slave. It will therefore result in a
type 01 failure response.
Master:
>99BB4[CR]
Slave:
N01[CR]

Acknowledge (C )

Requests a simple acknowledge from the slave. Useful to determine if
there is a device at that node address.
Master:
>99CB5[CR]
Slave:
A[CR]

Read Data (F)

The data field of this frame contains the handle for a system variable. See
Figures 5-7a and 5-7b for a list of defined handles for system variables.
:
Example
Master:
>34F0512[CR]
Slave:
A1234540[CR]