Figure A-4.
Setup tab.
A-6 | 5830 Moisture Analyzer
Setup Tab
Use the Setup tab to define analyzer parameters.
Gas
Select the gas being sampled.
Dewpoint Temp - If you select dewpoint temp, the Process Pressure box
is activated and you must enter a process pressure parameter in kPa.
NOTE
Units
Select the unit of measurement.
Gas Saver check box
Check this box to enable Gas Saver. This feature allows you to minimize
gas consumption.
4-20 mA Output
Set up your analog output range.
20 mA
Enter your high analog output limit.
4 mA
Enter your low analog output limit.
Hold during Zero
Check this box to hold analyzer output during verification
and Zero.
Alarm Output
Set up the limits for the concentration alarm.
Enable
Check this box to enable the concentration alarm.
High Limit
Enter the high limit for the concentration alarm.
Low Limit
Enter the low limit for the concentration alarm.
To Save settings on the Setup tab, click Apply.