In practice, some empirical modification is required; this helps to compen-
sate for the non-ideal performance of the meter. For example, the actual
factor for argon is 1.395, rather than the 1.401 estimated in the example.
Please consult AMETEK Process Instruments for assistance, before run-
ning gases that are not listed in the menu on the 5830 analyzer.
NOTE
Gas Saver
In the default configuration the Model 5830 analyzer runs with the in-
ternal bypass valve open, so that the total flow through the analyzer is
greater than 1 SLPM. This high sample flow rate helps to improve the
response speed of the system. The gas-saver feature allows the user to
control the state of the bypass valve. The operation of the valve can be
controlled by selecting either Enable (to close the bypass valve) or Disable
(to open the bypass valve) for the gas-saver option.

Clock

Use this function to set the clock on your analyzer.
Config
Clock
Year
Day
Day of
Week
Month
Hour
Minute
Do not enter a K-factor less than 0.25. A K-factor less than 0.25 will
cause erroneous flow and moisture generator readings. Contact Ame-
tek for support.
If a custom gas is entered, the conversion to units of dew point will be
based on an ideal gas model, without any enhancement factor.
Enter Value
Enter Value
Sunday
Monday
Tuesday
Wednesday
Enter Value
Thursday
Friday
Saturday
Enter Value
Enter Value
Year
Enter the 4-digit value for the year.
Day
Enter the 2-digit value for the day.
Day of the Week
Select the day of the week from the list.
Month
Enter the 2-digit value for the month.
Hour
Enter the value using the 24-hour clock.
Minute
Enter the 2-digit value for minutes.
|
User Interface
4-9