SPECIFICATIONS

All specifications subject to change without notice.
NOTE
Compatible Gases
High Pressure:
Air, argon, carbon dioxide, ethane, Freon 12, Freon 22, Freon 32, Freon 114,
Freon 116, Freon 125, Freon 134A, Freon 152, helium, hydrogen, kryp-
ton, methane, neon, nitrogen, oxygen, propane, SF6, and xenon. Carbon
dioxide requires a custom sensor assembly. Contact the factory to confirm
compatibility with other gases.
Low Pressure Compatible Gases:
Helium, Hydrogen, Nitrogen, Oxygen and Argon.
Read the Dryer Warning in the "Special Warnings and Information"
!
section in the beginning of this manual for more information.
CAUTION
Range
Calibrated range of 0 to 100 PPMV. Trend indication to 1000 PPMV
Limit of Detection
20 PPBV nominal
Accuracy
± 20 PPBV or ±10% of reading, whichever is greater
Response Time
Typical 63% of a step change in either direction in less than 5 minutes
Specifi cations | 2-1