Dryers; Removing The Dryers - Ametek 5800 User Manual

Moisture
Table of Contents

Advertisement

Figure 7-8.
Model 5830 dryers.
7-10 | 5830 Moisture Analyzer

Dryers

Removing the dryers

1.
At the rear of the analyzer, locate the dryer compartment on the right-
hand side (Figure 7-8).
Dryer Compartment
Cover
2.
Use a 7/64" hex wrench to loosen the four captured screws on the
compartment cover.
3.
Remove the cover from the analyzer.
4.
The sample system (reference) dryer is located in the top position, and
the zero dryer is located in the bottom position.
5.
Taking one dryer at a time, remove the mounting screws that pass
through the top of the dryer body and set them aside.
6.
Loosen the 1/8 inch VCR fittings on the dryer and remove the dryer
from the analyzer. Discard the used VCR gaskets from the fittings.
Sample System Dryer
(top dryer)
Zero Dryer
(bottom dryer)
Dryer Mounting Screws
(2 each)
1/8" VCR Gasket
(2 ea dryer)