Specifications; Utility Requirements; Outputs; Alarm Contacts - Ametek 5000 Manual

Moisture analyzer
Table of Contents

Advertisement

Specifications

For a complete listing of all analyzer specifications, refer to the analyzer User Manual.

Utility Requirements

All systems are available with choice of nominal AC voltages:
Field Unit:
• 100V ±10%, 50/60 Hz, 175W max.
• 115V ±10%, 50/60 Hz, 175W max.
• 230V ±10%, 50/60 Hz, 175W max.
Controller:
• 95–230V ±10%, 47-63 Hz, 75 W max.

Outputs

• 80-character vacuum fluorescent display
• Two 4 to 20 mA DC analog signals with completely user-defined span; one analog
output is auto-ranging
• RS-485 bidirectional serial port

Alarm Contacts

• System Alarm
• Concentration Alert
• Range or Verification Alert
(Contacts rated 30 VAC / 60 VDC, 50 VA maximum, 1A maximum, resistive)

Allowable Inlet Pressure

207 to 690 kPa gauge (30 to 100 PSIG)