Warranty
To obtain information about a warranty, if any, for this product, contact Covidien's Technical
Services Department, or your local Covidien representative (see back page).
Clinical Studies
Bland-Altman plots or SaO
vs error (SpO
- SaO
) are included to highlight the sampled data
2
2
2
points that were collected as part of the clinical evidence used to support the performance
specifications (SpO
accuracy) of this device.
2
1) This information can be especially useful to clinicians.
2) SpO
accuracy is validated by clinical tests, where the measured values of the sensors were compared with
2
those of the arterial co-oximetry in healthy adult subjects over the specified functional oxygen saturation
range. These values were validated on the following monitors: Nellcor™ PM1000N, N-600x, N-560, N-395. A
functional tester may not be used to validate SpO
accuracy.
2
3) Pulse rate accuracy was verified by simulated bench tests to ensure entire range was verified. These values
were verified on the following monitors: Nellcor™ PM1000N, N-600x, N-560, N-395.
7