Covidien Nellcor FLEXMAX Instructions For Use Manual page 4

Flexible spo 2 sensor, reusable
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
Note: Ensure the sensor is not incorrectly placed upside down (D), backwards (E), or with
the finger extending (F).
3. Switch on the oximeter and verify proper operation. (Ensure that the oximeter is providing
a reading.)
4. Visually monitor the sensor site at least every 6 hours to ensure integrity of the skin and
that no pressure injury is developing. Reposition to an alternate location if skin integrity
changes.
To remove the FLEXMAX, FLEXMAX-P model sensors:
1. Open the sensor by pressing on the sides of the sensor. Remove the sensor from the finger
and follow the cleaning instructions, if necessary.
2. Disconnect the sensor from the monitor by unplugging the sensor cable.
3. Store the sensor away for its next use.
Sensor expected life is 1 year.
WARNINGS
WARNING: Pulse oximeter probes are designed for use with specific monitoring systems.
Incompatible components can result in degraded performance and electromagnetic
incompatibility, or patient injury. The operator is responsible for checking compatibility prior
to use.
WARNING: A functional tester cannot be used to assess the accuracy of pulse oximeter
probes.
WARNING: Check the patient skin condition at least every 6 hours and reposition to an
alternate location if skin integrity changes.
WARNING: Misapplication of a pulse oximeter probe with excessive pressure and for
prolonged periods can cause pressure injury.
WARNING: Avoid application of the sensors to areas with poor skin integrity.
WARNING: Do not sterilize by irradiation, steam, or ethylene oxide – refer to cleaning and
disinfection instructions. Use of agents other than specified may damage the sensor.
WARNING: Do not use the sensor if it is damaged. Use of a damaged sensor could cause
patient injury or equipment failure.
WARNING: Excessive patient motion, excessive ambient light, electromagnetic interference
(e.g., stacking and location to other medical equipment), dysfunctional hemoglobin,
intravascular dyes, fingernail polish and long or artificial fingernails may affect the sensor
performance and the accuracy of the measurement.
WARNING: Route all cables carefully to reduce the possibility of patient entanglement or
strangulation.
WARNING: Do not alter or modify the sensor. Alterations or modifications may affect
performance or accuracy.
WARNING: Do not use the Nellcor™ sensor or other oximetry sensors during MRI scanning.
Conducted current may cause burns.
4