Language Dictionary

CALibrate:SECure:STATe

CALibration:SECure:STATe
This command enables and disables calibration mode. The calibration mode must
be enabled before the ac source will accept any other calibration commands. The
first parameter specifies the enabled or disabled state. The second parameter is
the password. It is required if the calibration mode is being enabled and the
existing password is not 0. If the password is not entered or is incorrect, an error is
generated and the calibration mode remains disabled. The query statement
returns only the state, not the password.
Whenever the calibration state is changed from enabled to disabled, any new
calibration constants are lost unless they have been stored with CALibrate:SAVE.
Unsecures or secures the ac source for calibration.
Returned parameters 0 | 1
66
Command syntax CALibrate:SECure:STATe <bool>[, <NRf>]
CALibration:SECure:STATe <bool>[, <NRf>]
Parameters 0 | 1 | OFF | ON [, <password>]
*RST value OFF
Examples
CAL:SEC:STAT 1, 6812 | CAL:SEC:STAT OFF
Query syntax CALibrate:SECure:STATe?
CALibration:SECure:STATe?
Related commands CAL:STAT
Keysight Models 6811C, 6812C, and 6813C Programming Guide