3
Language Dictionary
MEASure:VOLTage:ACDC?
FETCh:VOLTage:ACDC?
These queries return the AC+DC rms voltage being sourced at the output
terminals.
Returned parameters <NR3>
MEASure:VOLTage:HARMonic?
FETCh:VOLTage:HARMonic?
These queries return the rms amplitude of the Nth harmonic of output voltage.
The parameter is the desired harmonic number. Queries sent with a value of 0
return the DC component. A value of 1 returns the fundamental output frequency.
Harmonic orders can be queried up to the fundamental measurement bandwidth
of the measurement system, which is 12.6 kHz. Thus the maximum harmonic that
can be measured is dependent on the output frequency. Any harmonics that
represent frequencies greater than 12.6 kHz are returned as 0.
Returned parameters <NR3>
100
Query syntax MEASure:[SCALar]:VOLTage:ACDC?
FETCh:[SCALar]:VOLTage:ACDC?
Parameters None
Examples
MEAS:VOLT:ACDC?
Related commands MEAS:CURR:ACDC?
Query syntax MEASure:[SCALar]:VOLTage:HARMonic[:AMPLitude]? <NRf>
FETCh:[SCALar]:VOLTage:HARMonic[:AMPLitude]? <NRf>
Parameters 0 to 50
Examples
MEAS:VOLT:HARM? 3
Related commands MEAS:VOLT:HARM:PHAS?
Keysight Models 6811C, 6812C, and 6813C Programming Guide
FETC:VOLT:ACDC?
MEAS:VOLT?
FETC:VOLT:HARM? 1
MEAS:VOLT:HARM:THD?