CALibrate:PWM:RAMP

CALibration:PWM:RAMP
This command modulates the slope of voltage ramp driving the power output
stages. Varying the ramp affects the harmonic distortion of the output. The
argument is a number from 0 to 255. This command is only used during
manufacture or repair of the ac source.
Returned parameters <NR1>

CALibrate:SAVE

CALibration:SAVE
This command can only be used in calibration mode. It saves any new calibration
constants (after a current or voltage calibration procedure has been completed) in
non-volatile memory.
64
Command syntax CALCulate:PWM:RAMP <NRf>
CALibration:PWM:RAMP <NRf>
Parameters 0 through 255
Examples
CAL:PWM:RAMP 100
Query syntax CALibrate:PWM:RAMP?
CALibration:PWM:RAMP?
Related commands CAL:PWM:FREQ
Command syntax CALibrate:SAVE
CALibration:SAVE
Parameters None
Examples
CAL:SAVE
Related commands CAL:CURR
Keysight Models 6811C, 6812C, and 6813C Programming Guide
CAL:VOLT
CAL:STAT