Output:dfi; Output:dfi:source - Keysight Technologies 6811C Programming Manual

Ac power solutions
Hide thumbs Also See for 6811C:
Table of Contents

Advertisement

OUTPut:DFI

This command enables or disables the discrete fault indicator (DFI) signal to the
ac source.
Returned parameters 0 | 1

OUTPut:DFI:SOURce

This command selects the source for DFI events. The choices are as follows:
– QUEStionable: Questionable summary bit
– OPERation: Operation summary bit
– ESB: Standard Event summary bit
– RQS: Request Service summary bit
– OFF: Never true
Returned parameters <CRD>
Keysight Models 6811C, 6812C, and 6813C Programming Guide
Command syntax OUTPut:DFI[:STATe] <bool>
Parameters 0 | 1 | OFF | ON
*RST value OFF
Examples
OUTP:DFI 1
Query syntax OUTPut:DFI[:STATe]?
Related commands OUTP:DFI:SOUR
Command syntax OUTPut:DFI:SOURce <source>
Parameters QUEStionable | OPERation | ESB | RQS | OFF
*RST value OFF
Examples
OUTP:DFI:SOUR OPER
Query syntax OUTPut:DFI:SOURce?
Related commands OUTP:DFI
Language Dictionary
OUTP:DFI ON
3
105

Hide quick links: