MEASure:ARRay:CURRent:HARMonic:PHASe?

FETCh:ARRay:CURRent:HARMonic:PHASe?
These queries return an array of harmonic phases of output current in degrees,
referenced to the positive zero crossing of the fundamental component.
The first value returned is the DC component (always returned as 0 degrees
phase), the second value is the fundamental frequency, and so on up to the 50th
harmonic. Harmonic orders can be measured up to the fundamental
measurement bandwidth of the measurement system, which is 12.6 kHz. Thus the
maximum harmonic that can be measured is dependent on the output frequency.
Any harmonics that represent frequencies greater than 12.6 kHz are returned as 0.
Returned parameters 51 NR3 values
Keysight Models 6811C, 6812C, and 6813C Programming Guide
Query syntax MEASure:ARRay:CURRent:HARMonic:PHASe?
FETCh:ARRay:CURRent:HARMonic:PHASe?
Parameters None
Examples
MEAS:ARR:CURR:HARM:PHAS?
Related commands MEAS:ARR:VOLT:HARM:PHAS?