Voltage:offset:slew:mode; Voltage:offset:slew:triggered - Keysight Technologies 6811C Programming Manual

Ac power solutions
Hide thumbs Also See for 6811C:
Table of Contents

Advertisement

3
Language Dictionary

VOLTage:OFFSet:SLEW:MODE

This command determines how the DC offset slew rate is controlled during a
triggered output transient. The choices are as follows:
– FIXed: The offset slew rate is unaffected by a triggered output transient.
– STEP: The offset slew rate is programmed to the value set by
– PULSe: The offset slew rate is changed to the value set by
– LIST: The offset slew rate is controlled by the voltage list when a triggered
Returned parameters <CRD>

VOLTage:OFFSet:SLEW:TRIGgered

This command selects the DC offset slew rate that will be set during a triggered
step or pulse transient. A parameter of MAXimum or INFinity sets the slew to its
maximum possible rate. The SCPI representation for infinity is 9.9E37.
Returned parameters <NR3>
162
VOLTage:OFFSet:SLEW:TRIGgered when a triggered transient occurs.
VOLTage:OFFSet:SLEW:TRIGgered for a duration determined by the pulse
commands.
transient occurs.
Command syntax [SOURce:]VOLTage:OFFSet:SLEW:MODE <mode>
Parameters FIXed | STEP | PULSe | LIST
*RST value FIXed
Examples
VOLT:OFFS:SLEW:MODE STEP
Query syntax [SOURce:]VOLTage:OFFSet:SLEW:MODE?
Related commands VOLT:OFFS:SLEW
Command syntax [SOURce:]VOLTage:OFFSet:SLEW:TRIGgered <NRf+> | INFinity
Parameters 0 to 9.9E37 | MAXimum | MINimum | INFinity
Unit V (volts per second)
*RST value INFinity
Examples
VOLT:OFFS:SLEW:TRIG 50
Query syntax [SOURce:]VOLTage:OFFSet:SLEW:TRIGgered?
Related commands VOLT:OFFS:SLEW
Keysight Models 6811C, 6812C, and 6813C Programming Guide
VOLT:OFFS:SLEW:TRIG
VOLT:OFFS:SLEW:TRIG MAX
VOLT:OFFS:SLEW:MODE