3
Language Dictionary
SENSe:WINDow
This command sets the window function which is used in harmonic
measurements. KBESsel is the preferred window and should be used for most
measurements. RECTangular is available for making harmonic measurements that
comply with regulatory requirements for quasi-stationary harmonics.
When RECTangular is selected, the output frequency is constrained to frequencies
that give an integer number of cycles in the acquired waveform buffers, and the
measurement acquisition time is set to 0.1 seconds. Any programmed output
frequency will be routed to the closest frequency that has this attribute. These
frequencies are exact multiples of 10.000207 Hz.
Returned parameters <CRD>
116
Command syntax SENSe:WINDow[:TYPE] <type>
Parameters RECTangular | KBESsel
*RST value KBESsel
Examples
SENS:WIND KBES
Query syntax SENSe:WINDow?
Keysight Models 6811C, 6812C, and 6813C Programming Guide