3
Language Dictionary

VOLTage:MODE

This command determines how the AC rms output voltage is controlled during a
triggered output transient. The choices are as follows:
– FIXed: The voltage is unaffected by a triggered output transient.
– STEP: The voltage is programmed to the value set by VOLTage:TRIGgered
– PULSe: The voltage is changed to the value set by VOLTage:TRIGgered for a
– LIST: The voltage is controlled by the voltage list when a triggered transient
Returned parameters <CRD>
158
when a triggered transient occurs.
duration determined by the pulse commands.
occurs.
Command syntax [SOURce:]VOLTage:MODE <mode>
Parameters FIXed | STEP | PULSe | LIST
*RST value FIXed
Examples
VOLT:MODE FIX
Query syntax [SOURce:]VOLTage:MODE?
Related commands VOLT
Keysight Models 6811C, 6812C, and 6813C Programming Guide
VOLT:MODE:LIST
VOLT:TRIG