FREQuency:SLEW:MODE
This command determines how the frequency slew rate is controlled during a
triggered output transient. The choices are as follows:
– FIXed: The frequency slew rate is unaffected by a triggered output transient.
– STEP: The frequency slew rate is programmed to the value set by
– PULSe: The frequency slew rate is changed to the value set by
– LIST: The frequency slew rate is controlled by the frequency slew list when a
Returned parameters <CRD>
FREQuency:SLEW:TRIGgered
This command sets the rate at which frequency changes during a triggered output
transient. Instantaneous frequency changes can be obtained by sending
MAXimum or INFinity. The SCPI keyword INFinity is represented by the number
9.9E37.
Returned parameters <NR3>
Keysight Models 6811C, 6812C, and 6813C Programming Guide
FREQuency:SLEW:TRIGgered when a triggered transient occurs.
FREQuency:SLEW:TRIGgered for a duration determined by the pulse
commands.
triggered transient occurs.
Command syntax [SOURce:]FREQuency:SLEW:MODE <mode>
Parameters FIXed | STEP | PULSe | LIST
*RST value FIXed
Examples
FREQ:SLEW:IMM 75
Query syntax [SOURce:]FREQuency:SLEW:MODE?
Related commands FREQ
Command syntax [SOURce:]FREQuency:SLEW:TRIGgered <NRf+> | INFinity
Parameters 0 to 9.9E37 | MAXimum | MINimum | INFinity
*RST value MAXimum
Examples
FREQ:SLEW:TRIG 75
Query syntax [SOURce:]FREQuency:SLEW:TRIGgered?
Related commands FREQ
FREQ:SLEW MAX
FREQ:SLEW:TRIG
FREQ:SLEW:TRIG MAX
FREQ:SLEW:MODE