Error message
Component was con­
nected to another
device. Establish con­
nection?

17.2.4 Status signals

Battery charger connected
LED on
Status
power
LED on
supply
battery
charger
Battery charger disconnected
Beep
Vibration
signal
signal
1x short
1x short
96
Cause
The component was
connected to another
device/remote control.
A current connection to
the prosthesis was inter­
rupted.
Event
Power supply and battery charger operational. Battery charger not yet
connected to receiver.
Battery charger connected to receiver, good connection.
This display turns off automatically after approximately one minute to
avoid bothersome light at night. The charging process is not affected.
Event
Self-test completed successfully.
Product is operational.
Correction
To disconnect the original connection, tap
the "OK" button.
If the original connection is not to be dis­
connected, tap the "Cancel" button.
Check the following points:
Distance from the prosthesis to the
device
Battery charge level of the prosthesis
Bluetooth on the prosthesis switched
on? (see Page 79)
If multiple prostheses were stored in
the remote control, was the correct
prosthesis selected?
Resolution