15.2.3 Error messages while establishing a connection with the cockpit app

Error message
Device not found
Component was con­
nected to another
device. Establish con­
nection?

15.2.4 Status signals

Battery charger connected
LED on
Status
power
LED on
supply
battery
charger
46 | Ottobock
Charging status
indicator (5 LEDs)
Cause
The connection could
not be established as no
component with the
entered serial number
could be found.
The component was
connected to another
device/remote control.
A current connection to
the prosthesis was inter­
rupted.
Event
Power supply and battery charger operational. Battery charger not yet
connected to receiver.
Battery charger connected to receiver, good connection.
This display turns off automatically after approximately one minute to
avoid bothersome light at night. The charging process is not affected.
Error
The prosthesis is not
being charged
Defective battery
charger.
Correction
Compare the serial number entered with
that of the component, then try to establish
the connection again.
To disconnect the original connection, tap
the "OK" button.
If the original connection is not to be dis­
connected, tap the "Cancel" button.
Check the following points:
Distance from the prosthesis to the
device
Battery charge level of the prosthesis
Bluetooth on the prosthesis switched
on? (see Page 34)
If multiple prostheses were stored in
the remote control, was the correct
prosthesis selected?
Resolution
Reset the error by dis­
connecting and con­
necting the power sup­
ply.
If the error persists, the
battery charger and
power supply must be
inspected by an
authorised Ottobock
Service Centre.