Error message
NO NEW DEVICE
FOUND
14.2.4 Error messages while establishing a connection with the remote control app
Error message
DEVICE NOT FOUND
Component was con
nected to another
device. Establish con
nection?
14.2.5 Status signals
Battery charger is connected
LED on
LED on bat
power
tery charger
supply
Battery charger disconnected
Beep
Vibration
signal
signal
1 x short
1 x short
92 | Ottobock
Cause
No new prosthesis could
be found since this pros
thesis is already stored.
Cause
The connection could
not be established as no
component with the
entered serial number
could be found.
The component was
connected to another
device/remote control
A current connection to
the prosthesis was inter
rupted
Event
Power supply and battery charger operational
Event
Self-test completed successfully.
Correction
1) Confirm error message with the
2) Use the
,
keys to select the menu
option "SELECT DEVICE" and confirm
this with the key.
3) Use the
,
keys to select the pros
thesis you want and confirm this with
the
key.
Correction
Compare the serial number entered with
that of the component, then try to establish
the connection again.
To disconnect the original connection, tap
the "OK" button.
If the original connection is not to be dis
connected, tap the "Cancel" button.
Check the following points:
•
Distance from the prosthesis to the
device
•
Battery charge level of the prosthesis
•
Bluetooth on the prosthesis switched
on? (see Page 83)
•
Prosthesis switched on? (Switching off
the product)
•
If multiple prostheses were stored in
the remote control, was the correct
prosthesis selected?
key.