Error Messages While Establishing A Connection With The Remote Control App; Status Signals - Otto Bock C-Leg 3C98-3 Instructions For Use Manual

Hide thumbs Also See for C-Leg 3C98-3:
Table of Contents

Advertisement

Available languages

Available languages

Error message
NO NEW DEVICE
FOUND

14.2.4 Error messages while establishing a connection with the remote control app

Error message
DEVICE NOT FOUND
Component was con­
nected to another
device. Establish con­
nection?

14.2.5 Status signals

Battery charger is connected
LED on
LED on bat­
power
tery charger
supply
Battery charger disconnected
Beep
Vibration
signal
signal
1 x short
1 x short
92 | Ottobock
Cause
No new prosthesis could
be found since this pros­
thesis is already stored.
Cause
The connection could
not be established as no
component with the
entered serial number
could be found.
The component was
connected to another
device/remote control
A current connection to
the prosthesis was inter­
rupted
Event
Power supply and battery charger operational
Event
Self-test completed successfully.
Correction
1) Confirm error message with the
2) Use the
,
keys to select the menu
option "SELECT DEVICE" and confirm
this with the  key.
3) Use the
,
keys to select the pros­
thesis you want and confirm this with
the
key.
Correction
Compare the serial number entered with
that of the component, then try to establish
the connection again.
To disconnect the original connection, tap
the "OK" button.
If the original connection is not to be dis­
connected, tap the "Cancel" button.
Check the following points:
Distance from the prosthesis to the
device
Battery charge level of the prosthesis
Bluetooth on the prosthesis switched
on? (see Page 83)
Prosthesis switched on? (Switching off
the product)
If multiple prostheses were stored in
the remote control, was the correct
prosthesis selected?
key.