LED
LED on
Battery char
on
battery
ger connec
power
char
ted to product
sup
ger
ply
The
LED
turns off
or
change
s colour
at irreg
ular
inter
vals
Beep signal
4 x short at intervals
of approx. 20 sec.
(continuously)
14.2.3 Status signals
Battery charger connected
LED on
LED on
power
battery
supply
charger
Battery charger disconnected
Beep
Vibration
signal
signal
1 x short
1 x short
–
3 x short
90
Error
Yes
Temperature of the battery
charger too high
Error
Charging the battery outside the
allowable temperature range
Event
Power supply and battery charger operational
Event
Self-test completed successfully. Product is operational.
Maintenance note
Conduct the self-test again by connecting/disconnecting the battery
charger. If the beep signal is emitted again, you should visit the O&P
professional soon. If necessary, they will forward the product to an
authorised Ottobock Service Center.
The product can be used without restrictions. However, vibration signals
may not be generated.
Resolution
The distance between the
battery charger and the
receiver on the knee joint
must not exceed 1 mm. If this
distance is too great during
the charging process, the
magnetic surface of the bat
tery charger can heat up and
interrupt the charging pro
cess.
Take the battery charger off
the knee joint, disconnect it
from the power supply and let
it cool down. If the error
recurs, the battery charger
must be inspected by an
O&P professional.
Resolution
Check whether the specified ambi
ent conditions for charging the bat
tery are met (see page 84).