16.2.3 Error messages while establishing a connection with the cockpit app
Error message
Device not found
Component was connec
ted
to
another
device.
Establish connection?
16.2.4 Status signals
Battery charger connected
LED on
Status LED
power sup
on battery
ply
charger
Battery charger disconnected
Beep sig
Vibration
nal
signal
1x short
1x short
Genium 3B1/3B1=ST
Cause
The connection could not be
established as no component
with the entered serial num
ber could be found.
The component was connec
ted to another device/remote
control.
A current connection to the
prosthesis was interrupted.
Event
Power supply and battery charger operational. Battery charger not yet connected to
receiver.
Battery charger connected to receiver, good connection.
This display turns off automatically after approximately one minute to avoid bother
some light at night. The charging process is not affected.
Event
Self-test completed successfully. Product
is operational.
Correction
Compare the serial number entered with that of the
component, then try to establish the connection
again.
To disconnect the original connection, tap the "OK"
button.
If the original connection is not to be disconnected,
tap the "Cancel" button.
Check the following points:
•
Distance from the prosthesis to the device
•
Battery charge level of the prosthesis
•
Bluetooth on the prosthesis switched on? (see
Page 83)
•
If multiple prostheses were stored in the remote
control, was the correct prosthesis selected?
Resolution
Appendices