Frequency
Several times a day
/ according to local
practice and
standards
Every day / when
necessary
Every 2 weeks (at
least)
Every 6 months
( * )
Every year ( * )
6-8
Component
Patient circuit
Patients filter
Condensation trap
filter
Lung ventilator
Turbine air filter
Turbine air filter
Lung ventilator
Oxygen sensor
(optional)
Turbine air filter
Washers / O-Rings
Lung ventilator
Procedure / Action
Check for any water collection, drain and clean the
tubes when necessary.
Check for wear.
Check for any water collection, drain and clean
when necessary.
General cleaning and checks.
Replace. Components that cannot be destroyed
should be sterilized before disposal.
The lung ventilator must be inspected and
checked in general and any worn parts must be
replaced.
Use the appropriate preventive maintenance kit.
Replace. The working life of the cell depends on
the working environment. If the temperature or the
O
% is high, the working life of the sensor will be
2
lower.
Replace. Sterilize according to the procedure
described in this manual and according to local
standards.
Components that cannot be destroyed should be
sterilized before disposal.
Check the lung ventilator performance.
This includes an electrical safety test and
inspection of the ventilator for mechanical damage
and legibility of the labels.
The lung ventilator must also be inspected and
checked in general and worn parts must be
replaced, using the appropriate preventive
maintenance kit.
Use the appropriate preventive maintenance kit.