SIMV Breathing rate
SIMV Inspiratory time
Tidal volume (Vt)
Inspiratory pressure limit
( Pinsp - IPAP )
Inspiratory ramp slope
PEEP - EPAP
PEEP adjustment Microprocessor controlled electro-valves
O 2 concentration
Trigger detective method
Flow trigger ( I )
Trigger E
Inspiratory flow (FLOW)
Flow-by
PS (pressure support)
CPAP/PSV
Other controls
Patient circuit
Expandability
USER INTERFACE
Monitor
Dimensions 5,7"
Display keyboard
Falco 101
From 1 to 60 bpm
From 0.2 to 5.0 sec.
From 100 to 3000 ml (Adult)
From 50 to 400 ml (Paediatric)
From 2 to 80 cmH 2 O (in function of low and high pressure alarm set)
1, 2, 3, 4 (acceleration slope gradient) - (4 max. acceleration) (in operative
modes by pressure only)
From OFF, 2 to 30 cmH 2 O
O 2 - Air mixture enrichment by O 2 inlet at low pressure
Through sensor ( Flow )
Flow adjustable from 1 L/min to 9 L/min (step of 1 L/min)
From 5 to 90 % of the inspiratory flow peak
190 l/min
Automatic
From 2 to 60 cmH 2 O ( PSV, SIMV )
Pressure: from 3 to 50 cmH 2 O
MENU function
Function to select Curves, Parameters' Map displaying
Single hose Adult/Paediatric patient circuit 160 cm. (EXP valve on
circuit)
Single hose patient circuit 160 cm with mask for wisper (circuit with
leakage) - Optional
Upgradable software
Module with TFT LED display
Lateral keyboard for rapid access of functions and encoder knob for:
selection, set up and confirmation of physiological breathing parameters
selection and direct activation of functions
3
A-