Siare Falco 101 User Manual page 89

Lung ventilator, turbine-driven ventilation
Table of Contents

Advertisement

With optimal PSV, the breathing pattern can be standardised (by increasing the Vte and
reducing the respiratory rate) and the work of breathing can be reduced, improving the
respiratory exchange ratios.
In this mode the patient's work of breathing is assumed by the ventilator. Each breath
initiated by the patient (Tr. I activated) is supported by the ventilator, that sends a gas flow
inside the airway, at a certain pre-set pressure, called support pressure (PS).
Falco 101
If the patient does not trigger (spontaneous breathing during the apnea time
set in MAIN MENU - SETUP), the system activates the APNEA acoustic
and visual alarm.
The system will automatically provide an APCV ventilation with set safety
respiratory rate (RR bk) and I:E ratio (I:E bk).
4-17