4.4.1 Procedure for setting the PATIENT DATA

The fields for entering the name and surname have a limit of 14 characters.
The lower-case letters are placed after the upper-case letters.
To save the data inserted in each parameter (e.g. Name), you have to scroll through all 14
available character spaces.
Turn the encoder knob to select the first parameter you
want to set, Name.
Press the knob to activate the text insertion inside
Name box (the first character space appears).
Turn the knob to enter "the first letter" of the patient
Name (first the upper-case letters, then the lower-case
letters).
Press the knob to confirm the letter entered in the first
highlighted text box (the second character space
appears).
Turn the knob to enter "the second letter" of the patient
Name (first the upper-case letters, then the lower-case
letters).
Press the knob to confirm and turn it again to proceed
with entering the rest of the patient's Name.
Falco 101
You can see below how to enter the patient name (this is shown only as an
example).
Use the encoder knob to enter the patient's identification data.
Ventilator displaying the
PATIENT DATA screen.