Siare Falco 101 User Manual page 118

Lung ventilator, turbine-driven ventilation
Table of Contents

Advertisement

You choice a compensation factor and the you press the START button to confirm.
Wait the and of the calibration phase, then the device will begin the self-test. Press the
alarm when requested and the device enters in the STAND-BY mode.
To start the ventilation press the start key
Select Exit to pass to the self test screen and go
our from the " CALIBRATION PROGRAM "
screen.
4-46
To activate the inspiration flow sensor calibration program it is necessary the
intervention of SIARE authorized personnel or technical qualified personnel,
authorized by SIARE.
For additional information on the procedure please refer to the SERVICE manual.
If the user or the authorized technician don't active or choice any CALIBRATION
PROGRAM, after some seconds, the system goes automatically in SELF TEST
phase.
User Manual, DU3111103