" SELF TEST " : self-diagnostic tests list
Turbine functioning check and status.
During this test phase the system provide to 21%
calibration of the oxygen sensor.
Check the operation of the Inspiratory Flow sensor.
Check the operation of the pressure sensor by checking
PAW reading.
Expiratory electro-valve functioning check.
This test verify the patient circuit connection checking
the presence of pressure in the circuit opening and
closing the expiratory valve.
Check the battery voltage value.
Check the electric connection and the calibration of the
O 2 sensor (voltage delivered by the O
value between 9 to 14ml ).
The clinician should check if the system generates the
acoustic signal, he can confirm the test by silencing the
alarm.
3-16
During " SELF TEST " phase, the ventilator software carries out the
self-diagnostic tests and checks a series of devices necessary for safe
ventilator operation.
For a more correct and detailed analysis of the issues arising during " SELF
TEST " phase, please consult the Service Manual.
Turbine
Oxygen Emptying
Insp. Flow Sensor
Pressure Sensor
Electrovalve
Patient Circuit
Battery
Oxygen sensor
sensor at 21%:
2
Acoustic Alarm
User Manual, DU31111032