Set the low FiO
Set a FiO
The system activates the low FiO
Restore the default alarm value.
Set the high low FiO
Set a FiO
The system activates the high FiO
Restore the default alarm value.
During ventilation, set the main power supply switch to OFF ( 0 ).
The system activates the mains power failure alarm: silence
the alarm.
Restore the main switch and set it to ON ( I ).
3.7.5

Conclusions

3-28
concentration alarm limit to 50%.
2
concentration to 30% on the ventilator.
2
concentration alarm limit to 30%.
2
concentration to 50% on the ventilator.
2
Preliminary checks phase failed.
Please see Alarms chapter and/or Trouble Shooting chapter.
Please contact the nearest Siare Support Centre or any other support
centres authorised by Siare.
WARNING !! Patient injury hazard
Check the alarms set values before connecting a patient to the
ventilator.
Change the alarms setup based on the clinical situation.
WARNING !! Operator and patient injury hazard
The intensive care ventilator must be inspected and serviced once it
reaches 2000 hours of operation or, in case of limited use, at least once
every 6 months.
All maintenance and/or repair interventions require full knowledge of the
ventilator, and therefore such operations must be carried out only by
highly qualified staff, specifically trained and authorised by SIARE.
Any improper intervention or unauthorised modification may affect the
device's safety, putting the patient at risk.
alarm: silence the alarm.
2
alarm: silence the alarm.
2
LOW O2
CONCENTRATION
HIGH O
2
CONCENTRATION
POWER SUPPLY
FAULT
User Manual, DU31111032