Pinsp (cmH
O)
2
Maximum airway pressure limit value. The parameter is used in
pressure controlled modes to fix an operating limit for the airway
pressure that shall not be exceeded.
PMax (cmH
O)
2
Maximum airway pressure limit.
PS (cmH
O)
2
Positive airway support pressure value during inspiratory phase.
IPAP (cmH
O)
2
Airways pressure maximum value. This parameter is used in
BILEVEL ST mode and it's necessary to fix an operative limit
pressure which can't be passed.
Ti (s)
Time that defines the ventilator inspiration duration. The values
can be set based on the set RR.
Ti max (s)
Time that defines the maximum duration of an inspiration. If the
duration of the inspiratory phase is lower than the set value, the
patient will be forced to exhale.
Tr. E (%)
Percentage of the inhaled flow with regard to the maximum
peak where the inspiratory phase ends and the expiratory
phase begins.
Tr. I (L/min) (cmH
O)
2
Flow level (pressure) for detecting the patient spontaneous
breathing.
Vti (ml)
Inspired tidal volume guaranteed for each breath.