3.7.4 Alarms check

3-26
Check the alarms setup and if necessary change the values set,
based on the test you want to carry out.
• Turn the knob, select the alarm you want to change, press the knob to
activate the value editing function.
• Turn the knob clockwise to increase (anticlockwise to decrease) the
value of the selected alarm.
• Press the knob to confirm the modification of the alarm value.
Exit the ALARMS screen
• Press the ESC soft key; the system will leave the ALARMS screen.
• After about 60 seconds the system automatically returns to STAND-BY
or ventilator operation display.
WARNING !! Severe patient injuries
The alarms must trigger at the proper time and in the correct manner.
Check the proper activation of the visual and acoustic signals.
Before starting the alarm checking:
check the alarms setup, and change the set values whenever
necessary (please see previous paragraph)
Select pcv operative mode
Turn the knob: the system will
activate the ALARMS option.
Press the knob: the system will
activate the ALARMS page.
User Manual, DU31111032