Siare Falco 101 User Manual page 132

Lung ventilator, turbine-driven ventilation
Table of Contents

Advertisement

Soft keys and
encoder knob
(Patient) Circuit
disconnected
Low battery level
25% (50%)
Battery Autonomy
10 min.
5-14
This condition occurs when the control keyboard or the Encoder are
not working.
Switch
the lung ventilator OFF and then switch back ON.
If the problem persists, contact the Siare Service Centre or a
Centre authorized by Siare.
This alarm conditions occurs in case of disconnection of the patient
circuit (missed Vte detection for three times).
Check that the mask, endotrachéal tube and patient circuit are not
in some way split, disconnected or connected wrongly. If this is the
case, eliminate the problem or replace them.
Check the correct settings of the patient's respiratory parameters
(according to the operative mode selected: Volume/Flow, Rate, I/E,
Trigger).
Check that the patient circuit is connected correctly to the lung
ventilator and to the patient.
Contact the Siare Service
This alarm is activated when the charge level of the battery is at 25%
(50%) of the fully charged level.
Check that it is connected to the main power supply.
Recharge the battery.
If the alarm is activated when the battery has not provided the time
autonomy indicated on the technical sheet, request the intervention
of a Service Centre.
It is possible to silence the LOW BATTERY LEVEL 50% e LOW
BATTERY LEVEL 25% alarms pressing the ALARM RESET button on the
control keyboard.
This alarm is activated when the charge of the battery can grant an
autonomy of about 10 min.
Check the correct connection to the electric network: charge the
battery.
If the alarm occur when the battery doesn't delivery the autonomy
indicated in the technical sheet, kindly ask for a technical
intervention.
Centre or a Centre authorized by Siare.
User Manual, DU3111103