Soft keys and
encoder knob
(Patient) Circuit
disconnected
Low battery level
25% (50%)
Battery Autonomy
10 min.
5-14
This condition occurs when the control keyboard or the Encoder are
not working.
•
Switch
the lung ventilator OFF and then switch back ON.
If the problem persists, contact the Siare Service Centre or a
Centre authorized by Siare.
This alarm conditions occurs in case of disconnection of the patient
circuit (missed Vte detection for three times).
•
Check that the mask, endotrachéal tube and patient circuit are not
in some way split, disconnected or connected wrongly. If this is the
case, eliminate the problem or replace them.
•
Check the correct settings of the patient's respiratory parameters
(according to the operative mode selected: Volume/Flow, Rate, I/E,
Trigger).
•
Check that the patient circuit is connected correctly to the lung
ventilator and to the patient.
•
Contact the Siare Service
This alarm is activated when the charge level of the battery is at 25%
(50%) of the fully charged level.
•
Check that it is connected to the main power supply.
•
Recharge the battery.
•
If the alarm is activated when the battery has not provided the time
autonomy indicated on the technical sheet, request the intervention
of a Service Centre.
It is possible to silence the LOW BATTERY LEVEL 50% e LOW
BATTERY LEVEL 25% alarms pressing the ALARM RESET button on the
control keyboard.
This alarm is activated when the charge of the battery can grant an
autonomy of about 10 min.
•
Check the correct connection to the electric network: charge the
battery.
•
If the alarm occur when the battery doesn't delivery the autonomy
indicated in the technical sheet, kindly ask for a technical
intervention.
Centre or a Centre authorized by Siare.
User Manual, DU3111103