Ventilation phase

4.8
Press START soft key to begin the ventilation in the
selected mode with the most suitable PRP for the clinical
situation of the patient.
To returns in STAND-BY mode, please see 4.8.1.
In the upper section of the display you can find some values that will help you asses
the patient's clinical condition.
In the main section you can see the pressure curve.
4-28
Before starting the ventilation you have to:
carry out all ventilator hardware connections (medical gases, power supply,
patient circuit, ........)
carry out the preliminary checks (please see chapter 3.6)
set the language and the patient data
set and check the alarms limits
set the Physiological Respiratory Parameters and the operating mode that
match the patient's clinical situation best.
PRP
Based on the
set by the clinician according to the patient's characteristics,
the lung ventilator is able to display and measure a series of values necessary for
the patient's clinical evaluation.
User Manual, DU3111103