Falco 101
The value shows the Positive end of expiration Pressure: the
unit of measure is cmH
check if the ventilator joint the PEEP value level set.
The displayed value shows the real respiratory rate (number of
breaths per time unit) taking into consideration for the
calculation any spontaneous activity.
It shows the ratio between the inspiration time and the
expiration time.
It shows the oxygen concentration value (as percentage)
inhaled by the patient. The inhaled oxygen concentration value
is read by the system by means of the oxygen cell installed on
the inspiratory line (O2 sensor is optional)
It shows the current volume value during patient's inspiratory
phase: the unit of measurement is ml. The value is detected by
the flow sensor installed on the expiratory line.
It shows the volume value expired by the patient per minute :
the unit of measurement is L/min.
You can also calculate this value using the formula: current
volume (Vti) x respiratory rate (RR).
Description
O. Reading this value the user can
2
Vti : respiratory parameter
ml : unit of measurement
520 : measured value
2-11