A.4.2 Table 2 - Siare Falco 101 User Manual

Lung ventilator, turbine-driven ventilation
Table of Contents

Advertisement

A.4.2
Table 2
The device is intended for use in the electromagnetic environment specified below. The customer or
the user of the device should assure that it is used in such an environment.
Immunity test
Electrostatic discharge
(ESD)
EN 61000-4-2
Burst/Fast transient
EN 61000-4-4
Surge
EN 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
EN 61000-4-11
Power frequency
magnetic field
EN 61000-4-8
Falco 101
Immunity aspects
Test level
EN 60601-1-2
± 6 Kv contact
± 6 kV contact
± 8 kV air
± 8 kV air
± 2 kV power supply
± 2 kV power supply
lines
lines
±1 kV differential mode
±1 kV differential mode
±2 kV common mode
±2 kV common mode
<5 % U
<5 % U
T
(>95 % dip in U
)
(>95 % dip in U
T
for 0,5 cycles
for 0,5 cycles
40 % U
40 % U
T
(60 % dip in U
)
(60 % dip in U
T
for 5 cycles
for 5 cycles
70 % UT
70 % U
(30 % dip in UT)
(30 % dip in U
for 25 cycles
for 25 cycles
<5 % U
<5 % U
T
(>95 % dip in U
)
(>95 % dip in U
T
for 5 seconds
for 5 seconds
3 A/m
3 A/m
Electromagnetic environment -
Compliance level
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
T
)
T
Mains power quality should be that
of a typical commercial or hospital
T
)
environment. If the user of the
T
device requires continued operation
during power mains interruptions, it
T
is recommended that the device be
)
T
powered from an uninterruptible
power supply or a battery.
T
)
T
Magnetic power frequency fields
should be that of a typical
commercial or hospital
environment.
guidance
A-
15