The soft key, during the normal operating phase of ventilator,
allows the silencing of active acoustic alarm.
Pushing the RESET button will stop the acoustic alarm for a
defined time.
During the alarm silencing the alarm text is showed on the
panel.
Pushing further the RESET button will cancel the alarm text
only if the alarm conditions is disappeared.
If during the alarm silencing a new high priority alarm occur,
the alarm silencing is cancelled and the acoustic signal and
the visual texts are activated again.
Falco 101
It's possible to change the alarm settings even when the alarms are
activated.
After changing an alarm setting, the relevant sign is lighted and the status
icon will flash for a defined time.
Risk of injury for the user / patient
The user should never stop checking the patient conditions during the
alarm silencing.
5-7