Low voltage power supply
A proper connector for low voltage power supply
line (12Vdc / 7A) is placed on the ventilator rear
side.
Insert the 12Vdc power supply cable plug to
the plug placed on the back of the ventilator.
Battery power supply
3-6
WARNING !! Personal injury - physical hazard
In order to avoid any electric shock hazard, make sure that the supply
cable is connected to an electrical socket with the grounding cable
connected.
The lung ventilator complies with the requirements for electro-medical
devices detailed on chapter 1.5 ( Norms and standards regulations ).
To ensure proper operation of the lung ventilator, please connect to it
only additional devices that comply with the standards specified above.
The User should be always sure that the power supply electric plug is
accessible during the normal using and functioning of the device.
There should always be a battery installed inside the ventilator.
If there is no battery, the ventilator is not protected against voltage
drops or mains power supply cut off.
The ventilator should not be used without a charged battery.
The use with battery should be limited to short periods of time and
should not be considered an alternative to mains power supply.
Do not open the ventilator to replace the batteries or to carry out
maintenance operations on the same battery charger.
BATTERY LIFE
The battery operation time varies as follows: old battery or not fully
efficient, unusual ventilator parameters.
Replace the battery each two years (please refer to cap. 6 Maintenance).