A.4.3 Table 3 - Siare Falco 101 User Manual

Lung ventilator, turbine-driven ventilation
Table of Contents

Advertisement

A.4.3
Table 3
The device is intended for use in the electromagnetic environment specified below. The customer or
the user of the device should assure that it is used in such an environment.
Immunity test
EN 60601-1-2
RF conducted
EN 61000-4-6
RF Radiated
IEC 61000-4-3
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should be
less than the compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked with the following symbol:
A-16
Immunity aspects at R.F.
Test level
Compliance
level
3 Veff from
3 Veff from
150kHz to
150kHz to
80MHz
80MHz
3 Veff from
3 Veff from
80MHz to
80MHz to
2,5 GHz
2,5 GHz
Electromagnetic environment - guidance
Portable and mobile RF communications
equipment should be used no closer to any
part of the device, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency
of the transmitter.
Recommended separation distance
= 1,2 ⋅√P from 150kHz to 80MHz
d
d = 1,2 √P from 80 MHz to 800 MHz
d = 2,3 √P from 800 MHz to 2,5 GHz
where P is the maximum output power rating
of the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in metres
(m)
User Manual, DU3111103