Min. Vti / VM
Max. Vti / VM
Low (High)
PEEP/EPAP
PAW high
5-16
This alarm condition occurs in case the Vti is lower than set value.
Check that the corresponding alarm limits are set correctly.
Check the correct settings of the patient's respiratory parameters
(according to the operative mode selected: Volume/Flow, Rate,
I/E, Trigger).
Contact the Siare Service Centre or a Centre authorised by Siare.
This alarm condition occurs in case the Vti is higher than set value.
Check that the corresponding alarm limits are set correctly.
Check the correct settings of the patient's respiratory parameters
(according to the operative mode selected: Volume/Flow, Rate,
I/E, Trigger).
Contact the Siare Service Centre or a Centre authorised by Siare
This alarm condition occurs if the measured PEEP/EPAP value is
lower (higher) than the value set.
Check that the corresponding alarm limits are set correctly.
Check if the mask / endotracheal tube / patient circuit / expiratory
valve (membrane), are not damaged, if holes are present or if
they are connected in a bad way. If it's the case, replace them or
solve the trouble.
Check if the lung ventilator works properly verifying the airways
pressure. If the lung ventilator works properly check the flow
sensor calibration and its connection with the device.
In case of differences higher than 2 cmH2O (10%) between the
value set and the value read, a turbine calibration shall be
performed.
Contact the Siare Service Centre or a Centre authorised by Siare.
In this condition, the patient circuit + patient system presents a higher
resistance than expected or a lower compliance. This causes an
increase in airways pressure that exceeds the set limit.
Check that the corresponding alarm limits are set correctly.
Check that the mask, endotrachéal tube and patient circuit are not
in some way clogged, bent or crushed. If this is the case,
eliminate the problem or replace them.
Check the correct settings of the patient's respiratory parameters
(according to the operative mode selected: Volume/Flow, Rate,
I/E, Trigger).
Check that the luminous PAW bar on lung ventilator (the airways
pressure curve) correctly follows the inspiration / expiration cycle.
User Manual, DU3111103