Apnea
Turbine over-
temperature (over-
current)
Turbine failure
5-18
•
Check that the sensitivity of the Trigger is appropriate to the
patient's physiological conditions.
•
Check that the patient circuit is connected correctly to the lung
ventilator and to the patient.
•
If the problem persists, contact the Siare Service Centre or a
Centre authorised by Siare.
In this condition, no spontaneous respiratory activities is detected
(RR = 0).
•
Check that the patient's respiratory parameters are set correctly.
•
Check that the sensitivity of the Trigger is appropriate to the
patient's physiological conditions.
•
Check that the mask, endotracheal tube and patient circuit are not
in some way split, disconnected or connected wrongly. If this is
the case, eliminate the problem or replace them.
•
Check that the lung ventilator operates correctly, checking the
airways pressure trend. If the lung ventilator operates correctly,
check the flow sensor and the correct connection of its cable.
•
Check that the patient circuit is connected correctly to the lung
ventilator and to the patient.
•
Contact the Siare Service Centre or a Centre authorised by Siare.
The maximum safety level for the temperature (current) of the turbine
is passed (100 - 105 °C).
•
The lung ventilator switch off automatically to avoid dangers to the
patient safety.
•
If the problem persists, contact the Siare Service Centre or a
Centre authorised by Siare.
The turbine doesn't work properly (failure).
•
The lung ventilator switch off automatically to avoid dangers to the
patient safety.
•
If the problem persists, contact the Siare Service Centre or a
Centre authorised by Siare.