Pressure Graphic Display - Siare Falco 101 User Manual

Lung ventilator, turbine-driven ventilation
Table of Contents

Advertisement

2.5.4 Pressure graphic display

2-14
In case of alarm, the system displays the information below:
text string referring to the active alarm
"alarm bell" symbol indicating the alarm priority and status.
It is possible to modify the alarms settings ( MENU – ALARMS) even when
they are active.
After the modification, the relevant alert stays active and blinks for a
defined time.
ALARM SILENCING
Press the ALARM RESET key to interrupt the acoustic signal for a pre-
set period of time.
During the silencing period, the text of the alarm will still be displayed.
Press the ALARM RESET key once again to delete the alarm text, only
if the alarm activation condition is no longer present.
If during the silencing period, a new alarm (of high priority) occurs, the
alarms silencing command is automatically cancelled and the acoustic
and visual signals are reactivated.
WARNING!
Patient injury hazard
Alarms silencing. The clinician should not interrupt patient control during
alarms silencing period.
User Manual, DU3111103