Range of measured
parameters
Displayed parameters RR (bpm), I:E, ( * ) FiO 2 (%), Vti (ml), VM (L/min), PAW , PEEP,
Displayed graphics
Oxymeter
Calibration Automatic (started by the Operator)
ALARMS
Alarm types
Alarm priority
Alarms with limits set up by the operator
Airways pressure
Respiratory rate High – Low
Inspired tidal volume High – Low
Inspired minute volume High – Low
FiO 2 concentration High – Low
Electric power supply Alarm occurs in case of failure of external power supply
Apnoea Low Respiratory Rate (Apnoea BACK-UP function)
System alarms
Level (charge)
Level (charge)
Battery level (low)
Falco 101
•
Respiratory rate (range: 0 ÷ 200 bpm)
•
I:E ratio (range 1:99 ÷ 99:1)
•
( * ) % of O 2 (range: 0% ÷ 100%)
•
Tidal Volume: Vte, Vti (range: 0 ÷ 3000 ml)
•
Expired Minute Volume (range: 0 ÷ 40 l/min)
•
PAW: Pinsp - PS - IPAP, PEEP–EPAP-CPAP, (range: -20 ÷ 80cmH 2 O)
•
Inspiratory Peak Flow : Fi (range: 1 ÷ 190 l/min)
•
Expiratory Peak Flow : Fe (range: 1 ÷ 150 l/min )
•
Tinsp., Texp, Tpause (range 0.036 ÷ 10.9 sec)
CPAP (cmH 2 O)
•
CURVES: Pressure - Flow - Volume
•
Auto range
Electronic (value displayed in breathing parameters)
Optional : if FiO2 option is present
•
By MENU: with limits set by the operator
•
By DEFAULT: the operator cannot set them up
High - Mean - Standby
High – Low
Battery at 50%
Battery at 25%
10 Minutes
5
A-