3.6
Turn the lung ventilator OFF
•
STAND-BY operating mode.
•
Hold the ON-OFF key for few seconds to turn the
ventilator OFF.
Press NO: the ventilator goes back to
STAND-BY operating mode.
Press YES: the lung ventilator turns OFF.
3-20
CAUTION
Lung ventilator OFF - Oxygen emptying
During the switching OFF the ventilator performs a clean of the oxygen
inside the circuits and the device which grants a 21% concentration on
the inspiratory line of the lung ventilator.
This procedure grants: a longer oxygen sensor life and a quicker SELF
TEST phase when the device is turned-on.
The duration of this procedure depends by the FiO2 present in the
inspiratory line and can be of 60 seconds max.
Turn the lung ventilator OFF.