Set the low pressure alarm limit to 5 cmH
Disconnect the patient simulator from the patient circuit during
ventilation.
After about 20 seconds the system activates the airways low
pressure alarm: silence the alarm.
Reconnect the patient simulator.
Set the low respiratory rate (high respiratory rate) alarm limit to
15 bpm (15 bpm).
During ventilation please set RR = 10 bpm (RR = 20 bpm).
The system activates the low (high) respiratory rate alarm:
silence the alarm.
Restore the default alarm value.
Set the low alarm limit to Vt 250 ml.
During the ventilation, set a value to obtain a pressure value
lower than the value of the alarm.
The Low Inspiration Vt is activated : silence the alarm.
Set again the alarm to its default value.
Set the High alarm limit to Vt 500 ml.
During the ventilation, set a value to obtain a pressure value
higher than the value of the alarm.
The High Inspiration Vt is activated : silence the alarm.
Set again the alarm to its default value.
Falco 101
Press the START button: the
ventilator begins its cycle.
O.
2
LOW PRESSURE
LOW RESP. RATE
(HIGH RESP. RATE)
LOW INSP.
VOLUME
HIGH INSP.
VOLUME
3-27