4.11.2 Turbine characterization

Put the main switch of the ventilator (on the back
side of the device) in position n. "I".
Check on the ventilator keyboard if the green led
is turned on (presence of mains).
In order to access to the turbine characterization
(calibration), press and keep pressed for some
seconds the keys (in sequence), ALARM
RESET and ON-OFF .
The CALIBRATION screen is showed.
Connect the circuit to the simulator
When the screen is showeed press start to begin the turbine characterization
At the end of the calibration, the device starts the self test. Press the alarm reset key
when requested and the device goes in STAND-BY mode.
To start the ventilation press the start key