3.8.1 Turbine Characterization

The turbine characterization calibration is necessary in the following cases:
when you note differences out of tolerance (more than 2% of the end of the scale +
8% of the measured value) on the airways pressure values between the set and
measured values
software update or first turning-on after an important repairing (for ex. turbine
replacement or inspiratory flow sensor replacement)
3-36
In order to start the Turbine characterization (calibration) program is
necessary the intervention of qualified SIARE personnel or qualified
technical personnel authorized by SIARE.
Turn the encoder knob to
select the parameter to be
calibrated: the activated
message is highlighted.
Press the encoder knob: the
Turbine characterization
program is showed.