3.8.1 Turbine Characterization
The turbine characterization calibration is necessary in the following cases:
•
when you note differences out of tolerance (more than 2% of the end of the scale +
8% of the measured value) on the airways pressure values between the set and
measured values
•
software update or first turning-on after an important repairing (for ex. turbine
replacement or inspiratory flow sensor replacement)
3-36
In order to start the Turbine characterization (calibration) program is
necessary the intervention of qualified SIARE personnel or qualified
technical personnel authorized by SIARE.
•
Turn the encoder knob to
select the parameter to be
calibrated: the activated
message is highlighted.
•
Press the encoder knob: the
Turbine characterization
program is showed.
User manual, DU322210210